Abstract:Shot-to-shot fluctuations in laser-induced plasma emission limit the precision of quantitative laser-induced breakdown spectroscopy (LIBS). This study evaluates an ultrasonic-assisted normalization approach in which laser-ablation-induced body waves, measured with a contact piezoelectric transducer, are used as an external indicator of shot-dependent ablation variability. For each analyte, the first-arrival acoustic feature was related to the within-specimen relative deviation of the corresponding emission line using a single model fitted from pooled calibration data; the fitted coefficients were then applied unchanged to the prediction specimens. For Mn, Cr, and Cu, normalization reduced the shot-to-shot relative standard deviation (RSD) from 6.87%, 4.27%, and 4.59% to 2.30%, 2.25%, and 2.87%, respectively. The corresponding average relative errors decreased from 1.41% to 1.13% for Mn, from 2.27% to 2.02% for Cr, and from 1.04% to 0.93% for Cu. For comparison with spectral internal standardization, Fe-dominant reference features were selected from the averaged spectra using spectral non-overlap, monotonic calibration behavior, calibration-set R2 and RMSEC, and calibration-sample RSD; prediction metrics were not used during feature selection. The selected reference features were 391.843, 427.966, and 458.218 nm for Mn, Cr, and Cu, respectively. Both the optimized Fe-ratio and ultrasonic-normalized models improved quantitative performance relative to the raw-intensity models, although the extent of improvement depended on the analyte and evaluation metric. Additional measurements showed that the body-wave response is sensitive to mechanical coupling and specimen thickness, while a synchronous comparison with an airborne-acoustic channel showed distinct temporal and statistical characteristics. Accordingly, the contact body-wave approach can serve as a configuration-dependent external reference when stable mechanical coupling can be maintained.