Spectral Acquisition-Window Optimization for FL-LIBS Quantification of Cr, Mn, and Ni in Z3CN20-09M Steel
Author:
Affiliation:

Fund Project:

undefined

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
    Abstract:

    High-repetition-rate fiber-laser-induced breakdown spectroscopy (FL-LIBS) supports rapid surface scanning, yet reliable alloy quantification depends on spectra that are stable and representative of the sampled surface. Here, a spectral acquisition-window optimization strategy was developed for quantifying Cr, Mn, and Ni in Z3CN20-09M steel. Each acquisition window was characterized by pulse accumulation number and acquisition-window length, and CRITIC-TOPSIS integrated multiple spectral-quality metrics to rank the scanning conditions. Three representative acquisition windows were compared using full-spectrum partial least squares regression (PLSR) models calibrated with 11 high-alloy steel certified reference materials. The 10 ms, 4 mm s-1 condition gave the best overall performance, with a 40 μm acquisition-window length that approximately matched the laser spot diameter and a moderate pulse accumulation number of 350. Under this condition, full-spectrum PLSR yielded RMSEPs of 1.170, 0.063, and 1.975 wt.% for Cr, Mn, and Ni, respectively, lower than those under the matched comparison conditions. Subsequent model optimization using mutual information-variable importance in projection PLSR (MI-VIP-PLSR) further reduced the RMSEPs to 0.552, 0.047, and 0.127 wt.%. MI-VIP retained compact subsets of chemically informative variables, improving model interpretability while preserving high prediction accuracy. These results link pulse accumulation and spatial sampling to quantitative accuracy and provide a physically interpretable strategy for rapid, reliable FL-LIBS alloy quantification.

    Reference
    Related
    Cited by
Get Citation

Qi Yang, Jinna Mei, Huaiqing Qin, Yuhua Hang, Zhimin Lu, Shunchun Yao*. Spectral Acquisition-Window Optimization for FL-LIBS Quantification of Cr, Mn, and Ni in Z3CN20-09M Steel[J]. Atomic Spectroscopy,2026,47(5):777-792.

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:
  • Revised:
  • Adopted:
  • Online: October 07,2026
  • Published:
Copyright © 2026 Atomic Spectroscopy Press Ltd All rights reserved
Supported by:Beijing E-Tiller Technology Development Co., Ltd.