Abstract:Machine-learning performance in laser-induced breakdown spectroscopy (LIBS) is often summarized by a single accuracy, F1 score, R2 value, or root-mean-square error (RMSE), although the same metric can support very different claims depending on how the evaluation data were constructed and accessed. This work proposes a LIBS-specific evaluation-claim reporting framework that records physical independence, instrument/session separation, and evaluation access/custody separately from a factor-specific Domain-Shift Matrix (DSM) describing composition and concentration, physical state/preparation, ambient environment, excitation/geometry, detection/spectral response, and signal quality. Three public-data case studies illustrate the framework. A retrospective EMSLIBS 2019 reanalysis with public-label access examines whether previously reported internal-to-benchmark discrepancies recur under a different fixed protocol; its sample-disjoint variants also alter training composition. In NASA SuperLIBS quantification (997,625 spectra; 2,511 materials), random-shot validation underestimated matched-energy error on unseen in-range and compositionally out-of-distribution (OOD) targets by 0.185 and 0.735 standardized RMSE, whereas target- and split-group-disjoint validation reduced the corresponding absolute discrepancies to approximately 0.028 and 0.060. In a separate controlled lithology-classification study, training rows, labels, weights, preprocessing, and candidate trajectories were held fixed while only the validation view changed; population- and energy-aligned group-disjoint validation reduced median absolute validation-to-test gaps in all eight evaluated situations, but better agreement did not consistently improve locked-test accuracy. These empirical analyses are illustrative case studies rather than inferential estimates of the causal effect of validation design. The framework makes explicit what a reported score estimates and the conditions under which that interpretation is supported.